Product Features
*Discrete, random access, fully automated
*Up to constant 300 tests per hour
*Optional for external bar code reader
*120 positions for samples and 80 positions for reagents
*Automatic probe cleaning, liquid level detection, collision protection
*Reversed optic system with 8 wavelengths
*Refrigerated reagent and sample compartment
Product Parameters
Analysis system |
Test speed |
Constant 300T/h (non-ISE) |
Test principle |
Absorbance photometry, Turbidimetry |
Methodology |
End-point, Fixed-time, Kinetic, Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration |
Reaction plate |
81 reaction cuvettes |
Maximum reaction time |
15 minutes |
Minimum reaction volume |
150 ul |
Material of cuvettes |
Plastics (quartz glass optional) |
Linear range of absorbance: |
0-3.8Abs |
Light Source |
Halogen light |
Refrigeration system |
Water medium uniform refrigeration technology |
Wavelength |
340nm,405nm,450nm,510nm,546nm,578nm,630nm,670nm |
Reagent storehouse temperature |
4-16ºC |
Reaction temperature |
37±0.2ºC |
Temperature fluctuation |
±0.1ºC |
Sampling and mixing system |
Sample capacity |
2-30μl, 0.1μl increasing |
Reagent capacity |
2-300μl, 0.1μl increasing |
Reagent probe |
Sharing with the sample probe |
Reagent bottle |
15ml, 20ml, 30ml, 50ml, 60ml , 70ml can be used, the dead volume is less than 1ml |
Sample/reagent handling |
Sample/reagent tray |
120 positions for reagents and 80 positions for samples in refrigerated compartment (4~16ºC) |
Reagent volume |
R1:150~300μl, R2:20~150μl |
Sample volume |
2~100μl,step by 0.1μl |
Probe cleaning |
Interior and exterior automatic probe washing |
Calibration and quality control |
Quality control type |
real time quality control, daily quality control, day to day quality control |
The calibration type |
Linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method |
Operation system |
Software |
Windows2000, windows XP, windows7, windows8, windows10 |
Status display |
Real-time monitoring of sample tray,reagents tray,reacting tray; Real-time display of reaction temperature, reagent allowance, reaction curve, calibration curve and quality control chart |
FAQ
1 Q: Do you provide sample?
A: Sample will be provided when requested and the sample cost can be returned against your initial order.
2 Q: Do you accept Custom order?
A: Yes, we can help you custom your order, please reach out to us for more details.
3 Q: What about the payment methods for bulk order?
A: Generally we propose 30% deposit upon the order, the balance against the copy of B/L.
4 Q: What's your shipping solution?
A: Generally we send the order by air/train/truck (only for specific regions)/ocean, and by combination shipping such as 1st route by sea+2nd route by air/truck. It depends on your order volume and the principle is to make it affordable and favorable.
5 Q: Can you help take care of the clearance when the products arrives the local Customs?
A: We propose to do DDP terms and we can help you check if this service is available in your country.
6 Q: Can you consolidate the items and help me ship in one batch?
A: Sure, we can help you do that.
7 Q: How is your quality?
A: Quality comes from production, not in words. We can make sure the MP goods are the same as the sample you have approved.
8 Q: What's the best time to contact you?
A: Anytime you like, if urgent please reach out to us at WhatsApp.