Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Biochemistry Analyzer |
Still deciding? Get samples of US$ 4500/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Analysis system | ||
Test speed | Constant 300T/h (non-ISE) | |
Test principle | Absorbance photometry, Turbidimetry | |
Methodology | End-point, Fixed-time, Kinetic, Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration | |
Reaction plate | 81 reaction cuvettes | |
Maximum reaction time | 15 minutes | |
Minimum reaction volume | 150 ul | |
Material of cuvettes | Plastics (quartz glass optional) | |
Linear range of absorbance: | 0-3.8Abs | |
Light Source | Halogen light | |
Refrigeration system | Water medium uniform refrigeration technology | |
Wavelength | 340nm,405nm,450nm,510nm,546nm,578nm,630nm,670nm | |
Reagent storehouse temperature | 4-16ºC | |
Reaction temperature | 37±0.2ºC | |
Temperature fluctuation | ±0.1ºC | |
Sampling and mixing system | ||
Sample capacity | 2-30μl, 0.1μl increasing | |
Reagent capacity | 2-300μl, 0.1μl increasing | |
Reagent probe | Sharing with the sample probe | |
Reagent bottle | 15ml, 20ml, 30ml, 50ml, 60ml , 70ml can be used, the dead volume is less than 1ml | |
Sample/reagent handling | ||
Sample/reagent tray | 120 positions for reagents and 80 positions for samples in refrigerated compartment (4~16ºC) | |
Reagent volume | R1:150~300μl, R2:20~150μl | |
Sample volume | 2~100μl,step by 0.1μl | |
Probe cleaning | Interior and exterior automatic probe washing | |
Calibration and quality control | ||
Quality control type | real time quality control, daily quality control, day to day quality control | |
The calibration type | Linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method | |
Operation system | ||
Software | Windows2000, windows XP, windows7, windows8, windows10 | |
Status display | Real-time monitoring of sample tray,reagents tray,reacting tray; Real-time display of reaction temperature, reagent allowance, reaction curve, calibration curve and quality control chart |